Celltrion Recalls Dia Trust COVID-19 Ag Rapid Tests
3/17/22 — Celltrion USA is recalling specific lots of the DiaTrust COVID-19 Ag Rapid Test due to a high number of false positive reports. Additionally, the tests’ labeling for the affected products includes a shelf life of 18 months. However, the FDA’s emergency use authorization specifies these tests may only be used for 12 months. Read the FDA recall notice.